Bibliografie VÚGTK
Effects of impedance mismatch and coaxial cable length on absolute gravimeters
Křen, Petr ; Vaľko, Miloš ; Mašika, Pavel ; Pálinkáš, Vojtěch

Zdroj: Metrologia : 54(2), s. 161-170 ISSN: 1681-7575

Typ publikace: anotace
Rozsah: 10 stran

RIV: RIV/00025615:_____/17:N0000004; Fulltext: Scopus

Anotace:
Systematic effects of absolute gravimeters have to be investigated to fully utilize the capabilities of absolute gravimeters in metrology and geosciences. We found that for a FG5 gravimeter even a few meter long coaxial cables used for transmission of fringe signal cause systematic features in residuals and errors at the level of 1-2 µGal. In this paper we present experimental results and appropriate models for explanation of these effects that were found to be caused by impedance mismatches of electronic components and dispersion effects in coaxial cables of gravimeters.
Klíčová slova: absolute gravimeter ; laser interferometry ; dispersion ; impedance mismatch

Záznam se nachází v těchto sbírkách:
Výsledky VÚGTK > Vědečtí výzkumníci > Ing. Vojtěch Pálinkáš, Ph.D.
Výsledky VÚGTK > Podle útvarů / oddělení > 24: Geodézie a geodynamika
Výsledky VÚGTK > Vědečtí výzkumníci > Ing. Miloš Vaľko, Ph.D.
Dokumentační centrum VÚGTK > Články VÚGTK
Výsledky VÚGTK > Výsledky RIV
Anotace článků

 Záznam vytvořen 2017-06-02, poslední editace 2018-11-27



Hodnotit tento dokument:

Rate this document:
1
2
3
 
(Ještě nerecenzováno)