Bibliografie VÚGTK
On the effect of distortion and dispersion in fringe signal of the FG5 absolute gravimeters
Pálinkáš, Vojtěch

Source: Metrologia : 53(1), p.27-40 ISSN: 0026-1394

Publication type: anotace
Extent14 stran

RIV: RIV/00025615:_____/16:#0002201

Annotation:
The gravity reference, which results from the international comparisons held with the participation of numerous absolute gravimeters, is significantly affected by qualities of instruments prevailing in the comparisons (i.e., at present, FG5 gravimeters).Therefore, it is necessary to thoroughly investigate all instrumental (particularly systematic) errors. This paper deals with systematic errors of the FG5#215 coming from the distorted fringe signal and from the electronic dispersion at several electronic components including cables. In order to investigate these effects, we developed a new experimental system for acquiring and analyzing the data parallel to the FG5 built-in system.
Keywords: absolute gravimeter ; laser interferometry

Citation: PÁLINKÁŠ, Vojtěch. On the effect of distortion and dispersion in fringe signal of the FG5 absolute gravimeters. Metrologia. 2016, 53(1), 27-40. DOI: 10.1088/0026-1394/53/1/27. ISSN 0026-1394. Dostupné také z: http://iopscience.iop.org/article/10.1088/0026-1394/53/1/27
[Odkaz do RIVu]

The record appears in these collections:
Focus on VÚGTK > VÚGTK Departments > Geodesy and Geodynamics
Focus on VÚGTK > Researchers > Vojtěch Pálinkáš
Documents of VÚGTK > Articles VÚGTK
Focus on VÚGTK > RIV
Annotations

 Record created 2017-06-08, last modified 2019-05-20



Rate this document:

Rate this document:
1
2
3
 
(Not yet reviewed)